vialroom

#hplc-massspec 2026-05-23

Saturday23 messages6 participantstimes are UTC
Highlights from this day
  • factory_direct_fi — nobody runs enough blanks, including me 08:26
  • factory_direct_fi — baseline is drifting badly, is that the column or the mobile phase 08:58
  • bengaluru_bac — does ms/ms actually prove identity or just make it very likely 10:20
AA

coming back to this how much sample does a lab actually need for identity as well as purity

🤝5⚠️3
TM

ok so semaglutide is about 4113.6 Da. [M+2H]2+ lands near 2057.8, [M+3H]3+ near 1372.2
tirzepatide is about 4813.5 Da, so 2+ is near 2407.8 and 3+ near 1605.5

shallow gradients hide related substances behind the main peak. a steeper one separates and looks worse, honestly

DD

is monoisotopic or average mass the one on most certificates

BB

does ms/ms actually prove identity or just make it very likely

🤝1👀14🙏17
DD

a dimer usually shows up late and at roughly double the mass. the chromatogram alone will not tell you

right so two labs a point or two apart on the same lot is normal and gets mistaken for one of them being wrong constantly

a warm transit can put a visible degradant peak on a chromatogram. that is what 12 days at ambient does

MO

baseline drift on a gradient is normal. baseline drift on an isocratic hold is your column
a certificate with a number and no chromatogram is asking you to trust the integration you cannot see

MO

oxidation is about plus 15.99 Da and is usually methionine or tryptophan
214nm sees the amide backbone so it sees everything. 280 only sees aromatics

AA

unrelated but a certificate with a number and no chromatogram is asking you to trust the integration you cannot see

TM

i have had a lot where the mass was exactly right and the purity was 96.8. wrong question, right answer

why does integration change the number so much

why do two labs integrate the same peak differently