vialroom

#hplc-massspec 2025-12-08

Monday32 messages7 participantstimes are UTC
Highlights from this day
  • pbs_pip — why do two labs integrate the same peak differently 18:11
  • monoisotopic — update on the earlier thing a dimer usually shows up late and at roughly double the mass. the chromatogram alone will not tell you, ymmv 18:19
  • pbs_pip — i asked VendorInvestigate for the raw trace and they sent it. that is the right answer to that question monoisotopic and average diverge more the bigger the molecule.… 22:33
PP

why do two labs integrate the same peak differently

🤝11📉8
MO

update on the earlier thing why does integration change the number so much

two labs a point or two apart on the same lot is normal and gets mistaken for one of them being wrong constantly

MO

update on the earlier thing a dimer usually shows up late and at roughly double the mass. the chromatogram alone will not tell you, ymmv

PP

if somebody says they ran a mass spec and does not show you a spectrum, that is a claim, not a result

PP

i was wrong about tailing factor. it is usually the injection or the pH, not the column

baseline is drifting badly, is that the column or the mobile phase

[edited]

i have had a lot where the mass was exactly right and the purity was 96.8. wrong question, right answer

AB

has anyone ever had a result where the mass was right and the purity was wrong

QQ

relative retention time against a standard is comparable between labs. absolute minutes are not

AA

slightly off topic but is monoisotopic or average mass the one on most certificates

show me the spectrum

genuine question if the lab reports area percent without stating the wavelength, the number is less useful than it looks

how do you spot a deletion in a sequence from mass alone

QQ

truncations show as a mass deficit matching one or more residues. you need the sequence to say which

QQ

a certificate with a number and no chromatogram is asking you to trust the integration you cannot see

MO

the first peak is the solvent. it is always the solvent, take that with a pinch of salt

PP

baseline drift on a gradient is normal. baseline drift on an isocratic hold is your column

PP

214nm sees the amide backbone so it sees everything. 280 only sees aromatics

AA

update from 21 months ago: switched to asking for the chromatogram rather than the number and it changed everything

PP

tirzepatide is about 4813.5 Da, so 2+ is near 2407.8 and 3+ near 1605.5

TE

ms/ms sequencing is the only thing i would call identity. everything else is circumstantial

PP

i asked VendorInvestigate for the raw trace and they sent it. that is the right answer to that question
monoisotopic and average diverge more the bigger the molecule. say which one you are quoting

PP

nobody runs enough blanks, including me, i could be wrong