right so i have had a lot where the mass was exactly right and the purity was 96.8. wrong question, right answer
#hplc-massspec 2025-11-16
- vialkeeper — integration choices move the number by a few tenths easily. that is most of your inter-lab variance 17:44
- eat_more_please — deamidation is about plus 0.98 Da. it is not a rounding error, it is a degradant 18:50
- verify_or_dont — oxidation is about plus 15.99 Da and is usually methionine or tryptophan 19:17
- provincial_pat — how do you spot a deletion in a sequence from mass alone if the lab reports area percent without stating the wavelength, the number is less useful than it looks 20:29
need more sample
ms/ms or it isnt identity
whats a normal gradient for a peptide this size
is uv at 214 or 280 better for peptides
integration difference
nobody runs enough blanks, including me
baseline drift on a gradient is normal. baseline drift on an isocratic hold is your column, from memory
integration choices move the number by a few tenths easily. that is most of your inter-lab variance
retention time is a hint
do the arithmetic
within inter-lab range
if the number you are staring at is a third of what you expected, do the charge state arithmetic before you panic, ymmv
why do two labs integrate the same peak differently
update on the earlier thing relative retention time against a standard is comparable between labs. absolute minutes are not
thats co-elution
*that should say weekly
plus sixteen, is that always oxidation
you need more sample for identity than for purity. plan for that when you post it
ms/ms sequencing is the only thing i would call identity. everything else is circumstantial, n of 1 obviously
has anyone ever had a result where the mass was right and the purity was wrong
whats a plus one dalton shift usually
deamidation is about plus 0.98 Da. it is not a rounding error, it is a degradant
a dimer usually shows up late and at roughly double the mass. the chromatogram alone will not tell you
shallow gradients hide related substances behind the main peak. a steeper one separates and looks worse, honestly
show me the spectrum
i was wrong about tailing factor. it is usually the injection or the pH, not the column
[edited]shallow gradient
thats the solvent peak
a certificate with a number and no chromatogram is asking you to trust the integration you cannot see
oxidation is about plus 15.99 Da and is usually methionine or tryptophan
[edited]run a blank
plus 0.98 is deamidation
plus 16 is oxidation
if somebody says they ran a mass spec and does not show you a spectrum, that is a claim, not a result
how do you tell co-elution from a clean single peak
two labs a point or two apart on the same lot is normal and gets mistaken for one of them being wrong constantly
how do you spot a deletion in a sequence from mass alone
if the lab reports area percent without stating the wavelength, the number is less useful than it looks
is a tailing factor of anything specific a problem
thats a charge state
is there any point running a blank if the sample is the question
truncations show as a mass deficit matching one or more residues. you need the sequence to say which
the honest position is that hobby-level interpretation of a chromatogram is worth something but not much
the first peak is the solvent. it is always the solvent