  • non_scale_win — sorry to jump in i have had a lot where the mass was exactly right and the purity was 96.8. wrong question, right answer 17:58
  • bac_water_bill — if the lab reports area percent without stating the wavelength, the number is less useful than it looks, someone check my working 18:39
  • tenpence — i was wrong about tailing factor. it is usually the injection or the pH, not the column, take that with a pinch of salt 19:32

co-elution is the failure mode that a beautiful single peak is most likely to be hiding

plus 0.98 is deamidation

the first peak is the solvent. it is always the solvent, ill dig out the number

1
NS

quick one a certificate with a number and no chromatogram is asking you to trust the integration you cannot see

NS

sorry to jump in i have had a lot where the mass was exactly right and the purity was 96.8. wrong question, right answer

do the arithmetic

BW

if the lab reports area percent without stating the wavelength, the number is less useful than it looks, someone check my working

TE

if the number you are staring at is a third of what you expected, do the charge state arithmetic before you panic

BW

ok so truncations show as a mass deficit matching one or more residues. you need the sequence to say which

a shoulder on a peak is information, and usually bad information

TE

i was wrong about tailing factor. it is usually the injection or the pH, not the column, take that with a pinch of salt

19:35tenpence pinned a message to this channel
TE

ok so do you trust an area percent from a shallow gradient