#hplc-massspec 2025-09-02

Tuesday41 messages9 participantstimes are UTC
Highlights from this day
  • ms_ms_mira — then MS/MS if you actually want identity. fragment the parent ion and read the fragment ladder against the expected sequence. that is the real identity test and… 11:47
  • ms_ms_mira — they would not. that is why identity claims on new compounds are the thinnest paper in the whole market 12:41
  • vial_ledger — for context on what the molecule is meant to be 13:42
  • ms_ms_mira — ladder of evidence, weakest at the top 14:35
  • vial_ledger — we ran a mass spec is a sentence, not a result 15:02
AA

if somebody says they ran a mass spec and does not show you a spectrum, that is a claim, not a result

for the archive if the number you are staring at is a third of what you expected, do the charge state arithmetic before you panic

a certificate with a number and no chromatogram is asking you to trust the integration you cannot see

nobody runs enough blanks, including me

FT

if the lab reports area percent without stating the wavelength, the number is less useful than it looks

FT

214nm sees the amide backbone so it sees everything. 280 only sees aromatics

💀6🎉4🔥11
FT

while im here update from 10 months ago: switched to asking for the chromatogram rather than the number and it changed everything, thats one data point

AN

genuine question. can something be 99% pure and not be the peptide on the label

MM

yes, trivially, and that scenario is the one purity testing is worst at catching

VL

purity answers how much of this sample is one substance. identity answers which substance. different questions, different tests

MM

mass first. a mass spectrum with a sensible charge envelope tells you the molecular weight, and the wrong peptide is usually the wrong weight

MM

then MS/MS if you actually want identity. fragment the parent ion and read the fragment ladder against the expected sequence. that is the real identity test and almost nobody in this market pays for it

GN

there was a case in 2024 where an aggregated dataset showed a compound sold as retatrutide with a mass that did not fit anything in the family. purity fine, mass wrong

MM

and mass wrong plus purity high is the profile of a substitution, not a contamination. somebody made something cleanly, it just was not that

MM

they would not. that is why identity claims on new compounds are the thinnest paper in the whole market

😬11
VL

only for what the real molecule does, not for what is in your vial

VL

for context on what the molecule is meant to be

Cited study
Triple-Hormone-Receptor Agonist Retatrutide for Obesity — A Phase 2 Trial
New England Journal of Medicine · 2023
useful for what retatrutide is. tells you nothing about whether the powder you bought is it
FM

so what is the practical minimum before someone would trust an identity claim

MM

ladder of evidence, weakest at the top

EvidenceAnswers identity?Cost / availability
retention time vs standardweaklyincluded in most purity tests
co-injection with standardbetterneeds the lab to hold a standard
intact mass by LC-MSmostly, for gross errorssometimes offered
MS/MS sequence coverageyesrarely offered to hobbyists
amino acid analysispartly, composition onlyspecialist
a vendor saying it is genuinenofree
📌15😂6
GN

in fairness GGPeps put an intact mass figure on a sheet i got, which is more than the previous three vendors i used

MM

then it is a claim. a number typed on a page is not a spectrum. i am not saying they made it up, i am saying it is unverifiable

AN

Homopeptide sent me an actual spectrum image once. i could not read it but i appreciated it

MM

post it in here and someone will read it with you. that is half of what this channel is for

GN

and for what it is worth the JEEP sheets are purity-only and they do not pretend otherwise, which i respect more than a decorative mass figure