vialroom

#hplc-massspec 2025-07-08

Tuesday44 messages9 participantstimes are UTC
Highlights from this day
  • hk_hplc — how would a truncation sequence show up 13:52
  • hk_hplc — two labs a point or two apart on the same lot is normal and gets mistaken for one of them being wrong constantly 13:54
  • LC_MS_Lena — ms/ms sequencing is the only thing i would call identity. everything else is circumstantial 14:28
  • HPLC_Hank — the first peak is the solvent. it is always the solvent, i think 17:43
RR

sorry to jump in co-elution is the failure mode that a beautiful single peak is most likely to be hiding

HH

shallow gradients hide related substances behind the main peak. a steeper one separates and looks worse, honestly, still working it out

HH

how would a truncation sequence show up

😂9🧪9❤️13

two labs a point or two apart on the same lot is normal and gets mistaken for one of them being wrong constantly

👍2📉13😂2
RR

relative retention time against a standard is comparable between labs. absolute minutes are not, we shall see

ms/ms sequencing is the only thing i would call identity. everything else is circumstantial

💀6🤝15📉15

for the archive is monoisotopic or average mass the one on most certificates

[edited]
KF

i was wrong about tailing factor. it is usually the injection or the pH, not the column

[edited]
VB

Archive lookup: HPLC_Hank first appears in this channel on 2024-11-10.

HH

i have had a lot where the mass was exactly right and the purity was 96.8. wrong question, right answer

monoisotopic and average diverge more the bigger the molecule. say which one you are quoting

[edited]
HH

a certificate with a number and no chromatogram is asking you to trust the integration you cannot see

integration choices move the number by a few tenths easily. that is most of your inter-lab variance, i think
monoisotopic and average diverge more the bigger the molecule. say which one you are quoting

follow up do you trust an area percent from a shallow gradient

can you tell a dimer from the chromatogram alone

thats the solvent peak

TT

you need more sample for identity than for purity. plan for that when you post it

HH

if the lab reports area percent without stating the wavelength, the number is less useful than it looks

214nm sees the amide backbone so it sees everything. 280 only sees aromatics

do the arithmetic

the first peak is the solvent. it is always the solvent, i think

[edited]
1🙏11

thats a charge state

update from 3 months ago: switched to asking for the chromatogram rather than the number and it changed everything

*Medutest not the other one

quick one deamidation is about plus 0.98 Da. it is not a rounding error, it is a degradant