  • slow.taper — two labs a point or two apart on the same lot is normal and gets mistaken for one of them being wrong constantly 14:38
  • provincial_pat — truncations show as a mass deficit matching one or more residues. you need the sequence to say which 15:06
  • apnoea_down_two — if the number you are staring at is a third of what you expected, do the charge state arithmetic before you panic, n of 1 obviously 15:56
ST

two labs a point or two apart on the same lot is normal and gets mistaken for one of them being wrong constantly

🙏1

within inter-lab range

PP

truncations show as a mass deficit matching one or more residues. you need the sequence to say which

ST

you need more sample for identity than for purity. plan for that when you post it

AD

if the number you are staring at is a third of what you expected, do the charge state arithmetic before you panic, n of 1 obviously

shallow gradient

do you trust an area percent from a shallow gradient

AP

co-elution is the failure mode that a beautiful single peak is most likely to be hiding

ms/ms sequencing is the only thing i would call identity. everything else is circumstantial

plus 16 is oxidation

AP

oxidation is about plus 15.99 Da and is usually methionine or tryptophan